Appendix C: Data Port Protocol
Table 7: Voltage between pins 10 and 11
Alarm State
Nurse Call
Polarity Setting
Voltage from pins
10 to 11
No current alarms Normally high +5 to +12 VDC
Alarm condition Normally high -5 to -12 VDC
No current alarms Normally low -5 to -12 VDC
Alarm condition Normally low +5 to +12 VDC
Pins 7 and 15 provide a relay that closes when an alarm is
sounding on the monitor
. Pins 8 and 15 provide a relay that opens
when an alarm is sounding
. Pin 15 is a common lead for both
relays
. The relay operates whether the monitor is operating on AC
power or battery.
ANALOG OUTPUTS
The NPB-295 data port also provides analog voltage outputs
between pins 6, 13, 14, and ground (pin 10), which can be used to
calibrate the NPB-295 data port voltages with instruments such as
a chart recorder
. The voltage represents a specific measured
parameter’s current value
. The voltage differential varies
proportionally from 0 to 1 volt as the pin’s parameter varies over
its full range of values, as indicated in Table 8.
Table 8: Analog Pinouts
Pin Parameter Parameter Range
6 %SpO2 0 - 100%
13 Pulse rate 0 - 250 bpm
14 Pleth wave 0 - 255
For example, as the current value of %SpO2 varies from 0 to
100%, the voltage from pin 6 to ground (pin 10) would vary from
0 to 1 volt
. A voltage of .94 volts indicates a current %SpO2
value of 94.