Appendix C: Data Port Protocol
Software Revision Level
The next data field tells the user the software level, (Version
2.0.0.0) and a software verification number (CRC: XXXX)
.
Neither of these numbers should change during normal operation.
The numbers may change if the monitor is serviced and receives a
software upgrade.
Alarm Limits
The last data field in the top line indicates the high and the low
alarm limits for %SpO
2 and for the pulse rate (PR). In the
example above the low alarm limit for SpO
2 is 70% and the high
alarm limit is 100%
. Pulse Rate alarm limits are 60 and 160 bpm.
Column Headings
NPB-295 VERSION 2.0.0.0 CRC: XXXX SpO2 Limit: 70-100% PR Limit: 60-160BPM
TIME %SPO2 BPM PA Status
NPB-295 VERSION 2.0.0.0 CRC: XXXX SpO2 Limit: 70-100% PR Limit: 60-160BPM
TIME %SPO2 BPM PA Status
NPB-295 VERSION 2.0.0.0 CRC: XXXX SpO2 Limit: 70-100% PR Limit: 60-160BPM
TIME %SPO2 BPM PA Status
Actual column headings are in the second row of the column
heading line
. Patient data that is presented in the chart, from left
to right, is the time that the patient data on the row was obtained,
the current %SpO
2 value being measured, the current Pulse Rate
in beats per minute (BPM), the current Pulse Amplitude (PA),
and the operating status of the NPB-295.
Patient Data and Operating Status
Time
TIME %SPO2 BPM PA Status
12-NOV-98 14:00:05 100 120 50
The Time column represents the NPB-295 real-time clock.
75
Vedere la pagina 80