Mallinckrodt NPB-190 Manuale dell'Operatore Pagina 32

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 87
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 31
Start-Up and Use
Monitoring Mode
In Monitoring Mode - Pleth View (Figure 5), the NPB-295
displays SpO
2 readings, pulse rate readings, and a pleth
waveform. In the Monitoring Mode - Magnified (blip) View
(Figure 6), the Pulse Amplitude Indicator and a larger %SpO
2 and
pulse rate reading are displayed. The pleth waveform is not
displayed. How to select one of the two views by using the
softkeys will be discussed later in this section.
%SpO
2 is displayed for values between 0% and 100%. Pulse
rates are displayed for values from 20 to 250 beats per minute and
zero beats per minute. Pulse rates below 20 (except zero) will be
displayed as 20, and pulse rates above 250 will be displayed as
250. A pulse rate of zero is used to indicate that the monitor is
not monitoring a pulse.
A variable-pitch beep sounds once for each pulse, and the Pulse
Amplitude Indicator (in the Magnified View) visually displays
relative pulse strength at the sensor site. The pitch of the beep
decreases as %SpO
2 decreases.
Note: Verify that indicators, display information, and audible
sounds including alarms are operational, indicating that
the monitor is functioning. Each valid button push
should generate an appropriate audible or visual action.
Observe movement of the Pulse Amplitude Indicator or
pleth waveform, and listen for pulse beeps to verify that
measurements are being made.
If any action does not seem appropriate, do not use the
monitor. Instead, contact Mallinckrodt’s Technical
Services Department or your local Mallinckrodt
representative.
In Monitoring Mode, if the acquired pulse is lost, the monitor
enters Pulse Search Mode.
26
Vedere la pagina 31
1 2 ... 27 28 29 30 31 32 33 34 35 36 37 ... 86 87

Commenti su questo manuale

Nessun commento