
Principles of Operation
84
During monitoring, the instrument’s software reads this resistor
value and selects coefficients that are appropriate for the
wavelength of that individual sensor’s red LED; these
coefficients are then used to determine SpO
2. This resistor value
is read when the monitor is turned on, periodically thereafter, and
each time a new sensor is connected.
Additionally, to compensate for differences in tissue thickness,
the light intensity of the sensor’s LEDs is adjusted automatically.
Functional versus Fractional Saturation
This monitor measures functional saturation - oxygenated
hemoglobin expressed as a percentage of the hemoglobin that can
transport oxygen. It does not detect significant amounts of
dysfunctional hemoglobin, such as carboxyhemoglobin or
methemoglobin. In contrast, hemoximeters such as the IL482
report fractional saturation - oxygenated hemoglobin expressed
as a percentage of all measured hemoglobin, including measured
dysfunctional hemoglobins. To compare functional saturation
measurements to those from an instrument that measures
fractional saturation, fractional measurements must be converted
as follows:
functional saturation =
fractional saturation
100 - (% carboxyhemoglobin + % methemoglobin)
x100
Measured versus Calculated Saturation
When saturation is calculated from a blood gas partial pressure of
oxygen (PO2), the calculated value may differ from the SpO
2
measurement of a pulse oximeter. This usually occurs because
the calculated saturation was not appropriately corrected for the
effects of variables that shift the relationship between PO2 and
saturation (Figure 21): pH, temperature, the partial pressure of
carbon dioxide (PCO2), 2,3-DPG, and fetal hemoglobin.
Commenti su questo manuale